Autoclave

Performed validation to verify the Installational Qualification (IQ) and the Operational Qualification (OQ) of the Finn-Aqua Sterilizer.

This sterilizer is a custom built unit from Finn-Aqua specifically designed to meet cGMP and client's requirements. It is controlled by an external PLC (Allen Bradley) through an operator interface. Various cycle types can be programmed including wet and dry loads. All monitoring, controlling, and recording of the sterilization operation were performed via the PLC.

The validation verified the temperature distribution around the chamber via two pre-programmed cycles under loaded and unloaded conditions. Temperature distribution was monitored and recorded using a Kaye Digistrip.

Operational qualification included all controls, digital display, alarm functions, and cycle programs. Performance qualification included tests on thermal mapping, chamber leak test, DART Test (Daily Air Removal Test), and Heat Penetration Study.